Abstract

Detection of biomarkers at low concentrations is essential for early diagnosis of numerous diseases. In many sensitive assays, the target molecules are tagged using fluorescently labeled probes and captured using magnetic beads. Magnetic beads facilitate washing and separation steps, are well suited for automation, and improve the assay sensitivity. Current devices rely on quantifying the target molecules by detecting the fluorescence signal from individual beads. Thus, to detect low concentrations of target molecules, these devices require sophisticated optical detectors, making them bulky and expensive. Here, we propose a compact fluorescence-based system that simply uses a small permanent magnet with a conic tip to aggregate the magnetic beads, forming a cluster of fluorescently labeled probes whose fluorescence signal is much greater than that of a single bead. Using the magnetically aggregated biosensors to detect human Interleukin-8, we demonstrated a limit of detection of 0.1 ng/l and a 4-log dynamic range performance, which is on par with the most sensitive devices but is achieved without their bulk and cost.