Abstract
A method for preconcentration of Al(III) prior to its determination by inductively coupled plasma-optical emission spectrometry (ICP-OES) was developed because the sensitivity of ICP-OES is insufficient for trace-level measurement of Al(III). The method is based on the use of Coriolus versicolor}-immobilized $\gamma $-Fe$_{2}$O$_{3}$ magnetic nanoparticles as a solid phase sorbent. The best experimental conditions were found as pH 6.0, 2.0 mL/min flow rate, 100 mg amount of C. versicoloron 150 mg of $\gamma $-Fe$_{2}$O$_{3}$ nanoparticles, and 500 mL of sample volume. The Al(III) absorbed on biosorbent was eluted with 5.0 mL of 1.0 mol/L HCl. Limit of detection, limit of quantitation, linear range, correlation coefficient, relative standard deviation, and preconcentration factor were achieved as 0.03 ng/mL, 0.11 ng/mL, 0.2 to 20 ng/mL, 0.9994, 3%, and 100, respectively. Al(III) was preconcentrated in different water and food samples after successful results were achieved in the case of certified reference samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.