Abstract
A superconducting RF cavity has to be shielded from magnetic fields present during cool down below the critical temperature to avoid freezing in the magnetic flux at localized impurities, thereby degrading the cavity intrinsic quality factor Q <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">0</sub> . The magnetic shielding designed for the Fermilab vertical cavity test facility (VCTF), a facility for CW RF vertical testing of bare ILC 1.3 GHz 9-cell SRF cavities, was recently completed. For the magnetic shielding design, we used two cylindrical layers: a room temperature ldquoouterrdquo shield of Amumetal (80% Ni alloy), and a 2 K ldquoinnerrdquo shield of Cryoperm 10. The magnetic and mechanical design of the magnetic shielding and measurement of the remanent magnetic field inside the shielding are described.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.