Abstract
The self-locating peritoneal dialysis (PD) catheter, contains a tungsten tip. The effects of magnetic resonance (MR) on the catheter were evaluated, emphasizing its MR signal, artifacts, ferromagnetism, and possible heating production during the MR sequences. The catheter was studied in an ex vivo model using a 1.5T MR system and placed into a plastic box containing saline solution. Acquisitions on coronal and axial planes were obtained on fast gradient-echo T1-weighted and fast spin-echo T2-weighted. In vivo abdominal MR exams were also carried out. Overall, the catheter had good visibility. In all sequences, an extensive paramagnetic blooming artifact was detected at the level of the tip tungsten ballast, with a circular artifact of 5 cm in diameter. The catheter showed no magnetic deflection, rotation, or movements during all MR sequences. After imaging, the temperature of the saline solution did not change compared to the basal measurement. Patients safely underwent abdominal MR. The results point to the possibility of safely performing MR in PD patients carrying the self-locating catheter. The self-locating PD catheter is stable when subjected to a 1.5T MR system. However, it creates some visual interference, preventing an accurate study of the tissues surrounding the tungsten tip.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.