Abstract

BACKGROUNDThe resective surgery plus responsive neurostimulation (RNS) system is an effective treatment for patients with refractory focal epilepsy. Furthermore, the long-term intracranial electroencephalography data provided by the system can inform a future resection or ablation procedure. RNS patients may undergo 1.5-T magnetic resonance imaging (MRI) under the conditions specified in the RNS system MRI guidelines; however, it was unknown if the MRI artifact would limit intraoperative laser interstitial thermal therapy (LITT) in a patient with a fully implanted RNS system.OBSERVATIONSThe authors were able to complete a successful awake LITT of epileptogenic tissue in a 1.5-T MRI scanner on the ipsilateral side to an implanted RNS system.LESSONSIf a future LITT procedure is probable, the neurostimulator should be placed contralateral to the side of the potential ablation. Using twist drill holes versus burr holes for depth lead placement may assist in future laser bone anchor seating. Before a LITT procedure in a patient with the neurostimulator ipsilateral to the ablation, 1.5-T MRI thermography scanning should be scheduled preoperatively to assess artifact in the proposed ablation zone. Per the RNS system MRI guidelines, the patient must be positioned supine and awake, with no more than 30 minutes of active scan time before a 30-minute pause.