Abstract

Magnetic resonance imaging (MRI) is a powerful noninvasive diagnostic tool with superior soft tissue contrast. However, access to MRI is limited since current systems depend on homogeneous, high field strength main magnets (B0-fields), with strong switchable gradients which are expensive to install and maintain. In this work we propose a new approach to MRI where imaging is performed in an inhomogeneous field using radiofrequency spatial encoding, thereby eliminating the need for uniform B0-fields and conventional cylindrical gradient coils. The proposed technology uses an innovative data acquisition and reconstruction approach by integrating developments in field cycling, parallel imaging and non-Fourier based algebraic reconstruction. The scanner uses field cycling to image in an inhomogeneous B0-field; in this way magnetization is maximized during the high field polarization phase, and B0 inhomogeneity effects are minimized by using a low field during image acquisition. In addition to presenting the concept, this work provides experimental verification of a long-lived spin echo signal, spatially varying resolution, as well as both simulated and experimental 2D images. Our initial design creates an open MR system that can be installed in a patient examination table for body imaging (e.g., breast or liver) or built into a wall for weighted-spine imaging. The proposed system introduces a new class of inexpensive, open, silent MRIs that could be housed in doctor's offices much like ultrasound is today, making MRI more widely accessible.