Abstract

Airway stent implantation is a highly effective treatment for airway stenosis. However, it is presently unclear whether patients with airway stents can safely undergo magnetic resonance imaging (MRI). Such stents may be metallic or non-metallic, and MRI may induce stent dislodgment or heating and may be associated with stent-induced artifacts. We thoroughly reviewed the literature, experimental data, and manufacturer information on non-metallic, stainless steel (SS) and nickel-titanium alloy stents. Non-metallic stents are made of non-ferromagnetic materials associated with no MRI concerns. SS stents may shift in a magnetic field, generating significant artifacts. Nickel-titanium alloy stents are not at risk of dislodgement or heating, but may create some artifacts affecting image quality. Both non-metallic and nickel-titanium alloy stents are safe for patients who must undergo MRI. However, the safety of SS stents depends on the type of steel used.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call