Abstract

We conducted in vitro and in vivo experiments to study magnetic resonance imaging of intraocular foreign bodies. Diamagnetic and paramagnetic foreign bodies were imaged without artifact and without movement during the imaging process, while ferromagnetic foreign bodies produced large amounts of artifact that prevented meaningful images. All fetromagnetic foreign bodies moved during in vitro imaging. During in vivo imaging, three of four ferromagnetic bodies moved, producing substantial retinal injury. We concluded that magnetic resonance imaging is contraindicated in traumatized eyes with suspected ferromagnetic foreign bodies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.