Abstract

To determine the feasibility of performing in vivo magnetic resonance elastography (MRE) for uterine leiomyoma. Pilot study. Academic medical center. Six subjects planning surgical excision of uterine leiomyomas. MRE before planned surgery. Achieving an appropriate phase signal-to-noise ratio (PSNR) in the leiomyoma to allow assessment of leiomyoma elasticity in kilopascals (kPa). MRE was successful in all subjects for uteri ranging from 100 to >1,000 g. Subjects had body mass indexes ranging from 23.0 to 38.0 kg/m2. Appropriate PSNRs, ranging from 5.45 to 42.28, were achieved for leiomyomas in all subjects. Mean elasticity of uterine leiomyomas ranged from 3.95 to 6.68 kPa. MRE is a feasible technique for studying the in vivo mechanical properties of uterine leiomyomas and demonstrates significant heterogeneity in elasticity between lesions. Further work is necessary to optimize the technique and understand the clinical utility of this technique for women with uterine leiomyomas.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.