Abstract

Magnetic Resonance (MR) compatibility of biomedical implants and devices represents a challenge for designers and potential risks for users. This work addresses this problem and presents the first MR compatible multichannel silicon chronic microelectrode system, used for recording and electrical stimulation in the Central Nervous System of animal subjects. A standard chronic assembly was tested on a 2 Tesla magnet to detect areas of concern and modified to minimize or eliminate susceptibility artifacts, tissue damage, and electrode displacement, maintaining good image quality and safety to the animals. Multiple commercial connectors were tested for MR compatibility and several options for reference electrodes were also tested to minimize image artifacts. The required circuitry to interconnect the MR compatible microelectrode with standard data acquisition systems was also designed and fabricated. The final design is fully MR compatible and has been tested on guinea pigs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.