Abstract

Contrast-enhanced computed tomography angiography is usually valuable for the evaluation of clipped cerebral aneurysm, but it has side effects of contrast medium. Time-of-flight magnetic resonance angiography (MRA) is a non-invasive and fast method. However, clip-induced artifact limits assessment of the artery in the vicinity of a clip. MRA with ultrashort echo time (TE) reduces metal artifact. We use MAGNETOM Aera 1.5T (SIEMENS, München, Germany) and perform pointwise encoding time reduction with radial acquisition (PETRA)-MRA using ultrashort TE for the assessment of the cerebral aneurysm after clipping. We, herein, presented two representative cases with a clipped aneurysm which could be evaluated by PETRA- MRA. Especially in one of them, the neck remnant was revealed by PETRA-MRA. PETRA-MRA can reduce the time and the invasiveness and may be helpful for the usual follow-up of the clipped aneurysm with the development of MRA technology in the future.