Magnetic particle imaging (MPI) can be used as a fast diagnostic method for obtaining three-dimensional images from inside the body of small animals by the use of functionalized magnetic nanoparticles as tracer. Here, we present our scanner setup working at 10-kHz drive-field frequency to sample a field of view of 22 × 22 × 15 mm(3) with up to 32 volume images per second. A resolution of about 1 × 2 × 1 mm(3) is achieved with iron oxide nanoparticles (Resovist). We discuss the properties of the complete system for application in imaging small animals such as mice.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.