Abstract

Based on magnetic graphene (Fe3O4-G), a liquid chromatography with tandem mass spectrometric detection (LC-MS/MS) method coupled with magnetic solid-phase extraction was developed for the simultaneous detection of 47 kinds of nonsteroidal anti-inflammatory drug (NSAID) residues in meat of swine, chicken, and bovine. The samples were extracted with acetonitrile-phosphate, then the extracts were purified by dispersed solid-phase extraction based on magnetic graphene (Fe3O4-G). The linearity range is 0.1–50 μg L−1. Limits of detection were between 0.1 μg kg−1 and 0.5 μg kg−1, and limits of quantification were between 0.5 μg kg−1 and 1.0 μg kg−1. The average recoveries were 72.4–97.1% at spiking levels. The inter-day RSDs varied from 5.3 to 12.6%. Due to the advantages of easy collection of magnetic solid-phase extraction materials, the method was more rapid and efficient for the enrichment and separation without filling columns. The sensitivity, precision, and accuracy method was suitable for simultaneous detection of multi-component residues of nonsteroidal anti-inflammatory drugs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.