Abstract

Measurements were made of magnetic and electric field levels in and around a nuclear magnetic resonance imaging system undergoing a clinical trial. Magnetic field levels ranged from 0.04 tesla (T) in the imaging volume down to about 0.0006 T at the end of the patient table. The peak radio-frequency magnetic field level was 15 amperes per meter (A/m) in the imaging volume, while the rms value was 4.6 A/m. The specific absorption rate resulting from the radio-frequency magnetic field was calculated to be no more than 0.017 watts per kilogram (W/kg). The radio-frequency electric field was detectable only within a few centimeters of the coil assembly, and does not significantly contribute to the specific absorption rate. These exposure levels were much lower than existing guidelines for clinical NMR procedures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.