Abstract

Superconducting undulators and wigglers are developed for synchrotron light sources, damping rings for linear colliders and polarized positron sources. In an undulator the emitted photons along the trajectory have to interfere. In order to do so the magnetic field in all periods has to be almost identical. The field strength over one or several hundred periods is not allowed to deviate by more than 1%. Translated into mechanical accuracy the position of the wire and the poles has to be more accurate than about 5 over 1 to 2 m. High quality measurement of the field is an essential requirement. In this paper we present two field measuring systems, one is under construction and another one is under design phase at the Forschungszentrum Karlsruhe.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.