Abstract
In this work, we report on the sensitivity of photo-electrical detection of magnetic resonances (PDMR) for magnetometry measurement using low density nitrogen vacancy (NV) ensembles in CVD-grown diamond. We demonstrate that the selection of the laser excitation wavelength is of importance for achieving optimal magnetic field sensitivity. The PDMR sensitivity obtained using a yellow-green (561 nm) laser surpasses the performances of a green laser (532 nm), by suppressing the photoionization of defects other than NV centers (such as P1 centers). It consequently allows to carry out the PDMR measurements at lower laser powers with increased magnetic resonance contrast. Noticeably for both the green and the yellow-green illuminations, PDMR leads to an improved sensitivity to magnetic fields in the selected conditions compared to optically detected magnetic resonance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.