Abstract

The operation conditions of Gifford-McMahon (G-M) cryocoolers are limited by the external magnetic field induction. A G-M coldhead with the cooling power of 1.5 W at 4.2 K (RDK-415 by SHI) has been tested in an axisymmetric inhomogeneous magnetic field created by a conduction cooled superconducting magnet, which is able to generate up to 10 T. The tests revealed that the second-stage cooling power decreases by 29% at 1.0 T and by 42% at 1.6 T. The coldhead demonstrated abnormal operation in the magnetic field higher than some threshold value. The threshold field depends on the second-stage temperature. The stacks of YBCO tapes for the second-stage region and an iron shell for the motor region were tested as a magnetic shield. The magnetic field induction fluctuation on the refrigerator surface caused by periodical movement of magnetic regenerator in the inhomogeneous magnetic field is reported.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.