Abstract

Transcutaneous energy transmission (TET) is the most promising noninvasive method for supplying driving energy to a totally implantable artificial heart. Induction-heating (IH) cookers generate a magnetic flux, and if a cooker is operated near a transcutaneous transformer, the magnetic flux generated will link with its external and internal coils. This will affect the performance of the TET and the artificial heart system. In this paper, we present the design and development of a coil to be used for a magnetic immunity test, and we detail the investigation of the magnetic immunity of a transcutaneous transformer. The experimental coil, with five turns like a solenoid, was able to generate a uniform magnetic field in the necessary bandwidth. A magnetic-field immunity examination of the TET system was performed using this coil, and the system was confirmed to have sufficient immunity to the magnetic field generated as a result of the conventional operation of induction-heating cooker.

Highlights

  • Immediate energy transmission using a cable is an easy method for supplying driving energy to a totally implantable artificial heart [1,2,3,4,5]

  • We report on the design and development of a coil for the magnetic immunity test, and we detail the investigation of the immunity of transcutaneous transformers

  • We have reported on the design and development of a coil for the magnetic immunity test

Read more

Summary

Introduction

Immediate energy transmission using a cable is an easy method for supplying driving energy to a totally implantable artificial heart [1,2,3,4,5]. IH cookers generate a magnetic flux, and if a cooker is operated near a transcutaneous transformer, the magnetic flux generated will link with the transformer’s external and internal coils [7]. This can affect the performance of the TET system and the artificial heart system. We report on the design and development of a coil for the magnetic immunity test, and we detail the investigation of the immunity of transcutaneous transformers

TET System
Design of the Coil for the Magnetic Immunity Test
Magnetic-Field Immunity of the TET System
Conclusion