Abstract
To evaluate the effect of the static magnetic field of magnetic resonance (MR) scanners on keepers (ie, ferromagnetic stainless steel plate adhered to the abutment tooth of dental magnetic attachments). Magnetically induced displacement force and torque on keepers were measured using 1.5 Tesla (T) and 3.0 T MR scanners and a method outlined by American Society for Testing and Materials (ASTM). Changes in magnetic flux density before and after exposure to scanner static magnetic field were examined. The maximum magnetically induced displacement forces were calculated to be 10.3 × 10(-2) N at 1.5 T and 13.9 × 10(-2) N at 3.0 T on the cover surface. The maximum torques exerted on the keeper (4 mm in diameter) were 0.83 N × 4 mm at 1.5 T and 0.85 N × 4 mm at 3.0 T. These forces were considerably higher than the gravitational force (7.7 × 10(-4) N) of the keeper but considerably lower than the keeper-root cap proper adhesive force. The keepers' magnetic flux density remained less than that of the Earth. Magnetically induced displacement force and torque on the keeper in the MR scanner do not influence the keeper-root cap proper adhesive force.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.