Abstract
Introduction:Magnetically controlled growing rods (MCGRs) have been widely adopted in the management of early-onset scoliosis since they were first described in 2012. Recent reports have highlighted concerns around their safety. To date, little is understood about the risk factors and modes of failure in these devices.Case Report:We report a novel mechanism of device failure in a 14-year-old patient following multiple revisions of MCGRs.Clinically, there was no evidence of device failure and the MCGRs appeared radiologically intact. Explantation analysis revealed multiple compromised/non-functional components. A previously undocumented phenomenon of complete magnet fracture was also seen.Conclusion:The absence of clinical or radiological features of device failure in this case makes the findings of great concern. Given the relative paucity of high-quality evidence surrounding the use of MCGRs, we support calls for urgent comparative studies and further investigation of risk factors for device failure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.