Abstract

Magnamosis creates a secure sutureless anastomosis through magnetic compression. In this study, we further develop and test delivery devices capable of creating a secure duodeno-colonic anastomosis using available minimally invasive endoscopic techniques. Eight pigs underwent general anesthesia. Colonoscopy was used to deliver 1 magnetic ring to the hepatic flexure. Simultaneously, upper endoscopy delivered the other magnetic ring into the duodenum using a variety of techniques. The 2 magnetic rings were brought into magnetic proximity under laparoscopic guidance. The pigs were recovered and examined daily followed by sacrificing at 1, 2, 4, and 6 weeks. The device designed to deliver and release each magnetic ring evolved from using a guide wire and balloon devices to redesigning the magnetic ring casing with a groove to accommodate an endoscopic snare. Laparoscopic visualization assured safe magnet mating of intestinal segments. The duodeno-colonic anastomoses created with the snare yielded widely patent anastomoses. In vitro testing revealed excellent burst pressure. Histology revealed complete healing as early as 1 week. We redesigned the magnamosis device to facilitate delivery by endoscopic techniques. The snare technique allows endoscopic positioning and controlled release of the magnetic rings for a secure side-to-side duodeno-colonic anastomosis.