Abstract

Polycyclic aromatic hydrocarbons and polychlorinated biphenyls are commonly categorized as persistent organic pollutants. In order to analyze these pollutants, customized stationary phases are increasingly being developed and synthesized for solid-phase extraction. In this work, we tested a new solventless solid-phase extraction approach based on the use of a Magic Chemisorber® (Frontier Lab) which consists of a bead-covered polydimethylsiloxane stationary phase with a thickness of 500 µm. These devices are directly immersed into aqueous samples and then introduced into a pyrolysis–gas chromatography-mass spectrometry system equipped with a cryofocusing system for the thermal desorption and analysis of the adsorbed species. Our new method performs better than the most recent solid-phase extraction devices, with limits of detection lower than 2.7 ng/L and limits of quantification lower than 9.0 ng/L. The method was tested on standard compounds and on an environmental sample, showing the potential to characterize other chemical species besides the persistent organic pollutants, such as phthalate plasticizers and antioxidants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.