Abstract

Measurement of macular pigment (MP) can be performed by analysis of autofluorescence (AF) images. These can be obtained by standard 488-nm argon-imaging alone (one wavelength, 1-Lambda) or with additional digital subtraction of a second image at 514 nm (two wavelengths, 2-Lambda). The analyses are easy to perform, and we present a comparison of both methods and investigate their reliability and repeatability. Inter-individual variability of MP optical density (MPOD) measurements was assessed in single eyes of 120 subjects with a modified Heidelberg retina angiograph (HRA). MPOD values obtained with one (488 nm) Lambda (MPOD(1Lambda)) were compared with those obtained with two (488 nm and 514 nm) Lambda (MPOD(2Lambda)). To test the repeatability of the two methods, 20 subjects were subjected to five repeated measurements. Among 120 individuals, mean MPOD(1Lambda) at 0.5 degrees eccentricity was 0.59 (range 0.06-1.32), mean MPOD(2Lambda) was 0.5 (range 0.01-1.21). Apart from this systematic difference, 1-Lambda and 2-Lambda measurements at 0.5 degrees agreed well across the range of MPOD values (beta=0.964, around the fovea, a systematic difference (0.11) was accompanied by declining agreement at higher MPOD values (beta=0.669, 95% CI 0.519-0.844; R=0.48). Among 20 subjects with five repeated measurements, the reliability ratio was 0.97 for 1-Lambda and 0.94 for 2-Lambda at 0.5 degrees and 0.93 and 0.94, respectively, at a distance of 2 degrees. Both methods showed a high repeatability with little influence of measurement error. They agree well at the fovea centre in terms of ranking individuals according to their MPOD, but provide increasingly deviating results at a distance of 2 degrees around the fovea, probably because the 1-Lambda method, in contrast to the 2-Lambda method, cannot compensate for disruptive influences and for heterogeneous distributions of the lipofuscin fluorophores. The 1-Lambda method can be performed by standard HRA and could therefore be used for screening in multicentre studies, but only approaches the actual amounts of MP. The 2-Lambda method remains the more precise method for MPOD measurement in autofluorescence imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.