Abstract

The influence of the conditions of preparation of divinylbenzene-based polymer monoliths on the properties of monolithic capillary columns for use in gas adsorption chromatography was examined. It was found that the polymerization time and the temperature and composition of the polymerization mixture have an effect on the dynamic and chromatographic properties of the columns. The monolithic capillary columns prepared under the optimal synthesis conditions had the height equivalent of a theoretical plate at the level of 20–30 μm, which is an order of magnitude below that of conventional open tubular columns.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.