Abstract

The aim of the present work was to functionalize poly(vinyl chloride) (PVC) urinary catheters with grafted copolymers that can improve the biocompatibility and serve as binding points of lysozyme. PVC catheters were modified by grafting a mixture of N-vinylcaprolactam (NVCL) and 2-hydroxyethylmethacrylate (HEMA) applying a gamma-ray pre-irradiation method. The effect of absorbed dose, monomer concentration, temperature, and reaction time on the grafting percentage was evaluated. The grafted catheters were characterized regarding surface composition (FTIR-ATR spectroscopy), thermal properties (DSC and TGA) and swelling in aqueous medium. Lysozyme was directly coupled onto PVC-g-(NVCL/HEMA) previously activated using carbonyldiimidazole. Antimicrobial lytic activity of the modified catheters over time was tested against Micrococcus lysodeikticus. Lysozyme diminished the adhesion of Staphylococcus aureus onto the functionalized catheters, which may be suitable to prevent biofilm formation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.