Abstract

To develop 3D ultrashort-TE (UTE) sequences with tight TE intervals (δTE), allowing for accurate mapping of lungs under free breathing. We have implemented a four-echo UTE sequence with δTE (< 0.5 ms). A Monte-Carlo simulation was performed to identify an optimal number of echoes that would result in a significant improvement in the accuracy of the fit within an acceptable scan time. A validation study was conducted on a phantom with known short values (< 5 ms). The scanning protocol included a combination of a standard multi-echo UTE with six echoes (2.2-ms intervals) and a new four-echo UTE (TE < 2 ms) with tight TE intervals δTE. The human imaging was performed at 3 T on 6 adult volunteers. mapping was performed with mono-exponential and bi-exponential models. The simulation for the proposed 10-echo acquisition predicted over 2-fold improvement in the accuracy of estimating the short compared with the regular six-echo acquisition. In the phantom study, the was measured up to three times more accurately compared with standard six-echo UTE. In human lungs, maps were successfully obtained from 10 echoes, yielding average values = 1.62 ± 0.48 ms for mono-exponential and = 1.00 ± 0.53 ms for bi-exponential models. A UTE sequence using δTE was implemented and validated on short phantoms. The sequence was successfully applied for lung imaging; the bi-exponential signal model fit for human lung imaging may provide valuable insights into the diseased human lungs.