Abstract
Endotracheal misdirection of narrow bore nasogastric feeding tubes resulted in perforation of the lung, pneumothorax and hydrothorax in two intensive care patients. Both were intubated with cuffed endotracheal low pressure tubes, one patient was on respirator therapy with neuromuscular relaxation. Feeding tubes were inserted by experienced personnel with the assistance of a steel stylet without difficulties. Aspiration of fluid was misinterpreted as proof of correct positioning, the liquid being however pleural effusion and not gastric juice. Similarly auscultation of gurgling sounds in the upper epigastrium was not a reliable sign of intragastric position. Insertion of nasoenteric feeding tubes may be complicated by perforation of the upper gastrointestinal tract and lung in poorly responsive pateints with cuffed endotracheal devices during neuromuscular blockage. In these patients a laryngoscope and forceps should be used to ensure free passage of the tube into the oesophagus. Röntgenographic confirmation of correct positioning of the tube immediately after insertion is mandatory.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.