Abstract
A function has been developed that will fit counting efficiency to chest wall thickness, adipose content of the chest wall, and photon energy. The adipose content of the chest wall is removed as a variable by use of the derived quantity: muscle equivalent chest wall thickness. The function has been tested on experimental data sets obtained from the LLNL and JAERI torso phantoms using a variety of germanium detector lung counting systems. The function is applicable over the range of existing phantoms and the range of equipment types tested. Subtle differences between the detector systems are shown by differences in the fit parameters. This analysis could be a useful addition to lung counting analysis software. The LLNL and JAERI phantoms have very similar counting characteristics as evidenced by the similarity of the function parameters when measured using a single detector system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.