Abstract
To test the feasibility of a system for intermittent irradiation synchronized with respiratory motion in a clinical setting. A newly developed gate pulse controller that starts and stops irradiation at a chosen phase of the respiratory cycle by controlling a linear accelerator was used in six patients with lung cancer. A laser displacement sensor was used for the detection of respiratory motion. Three patients underwent radiation therapy during the cycles between 50% expiration and 50% inspiration (step 1), and three patients underwent radiation therapy during the cycles between 70% expiration and 30% inspiration (step 2). The system functioned well; irradiation was verified with portal verification radiography in all six patients. The range of the tumor position during synchronized irradiation was detectable with fast portal localization radiography. The treatment times for steps 1 and 2 were 1.38-1.71 and 2.03-2.18 times longer, respectively, than those for conventional irradiation. Synchronized irradiation with the authors' system allowed convenient and reliable reduction of the target volume. Further study is needed to standardize the system for clinical use.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.