Abstract
Abstract In the framework of Total Hip Replacement (THR), the aim of this paper was to propose a novel 3D contact-lubrication model for the in-silico prosthetic wear calculation during the gait. The approach is based on the Reynolds equation and in-vivo loads were selected for the simulations. The hip relative angular velocities were calculated by a commercial musculoskeletal software. In the considered conditions, by using the Archard wear theory, the model allows the determination of the unsteady fluid film pressure and the volumetric wear rate prediction. With reference to a Ceramic on Polyethylene hip prosthesis, the volumetric wear rate was calculated showing a good agreement with the values obtained by other Authors as a result of FEM approaches or in-vitro tests.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.