Abstract
Two topics on our recent investigations into superconducting tunnel junction (STJ) detectors are discussed: the single Nb/Al-AlOx/Al/Nb junction for high-resolution x-ray detection and the series-connected superconducting tunnel junction detectors (SCSDs) with high detection efficiency. Using a single Nb/Al-AlOx/Al/Nb junction with rather large area of 178 X 178 micrometers 2, we obtained a high energy resolution of 88 eV for 5.9-keV x- rays. The signal-to-noise ratio of the SCSD is discussed from a simple theoretical formulation of electronics, showing that the series-connected STJs can be developed as a nuclear radiation detector both with high energy resolution and high efficiency. Recent experimental results of our SCSD are also given in this paper.© (1992) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.