Abstract

We have developed a weighted optical flow algorithm for the extraction of instantaneous blood velocity from dynamic digital x-ray images of blood vessels. We have carried out in- vitro validation of this technique. A pulsatile physiological blood flow circuit was constructed using sections of silicone tubing to simulate blood vessels with whole blood as the fluid. Instantaneous recording of flow from an electromagnetic flow meter (EMF) provided the gold standard measurement. Biplanar dynamic digital x-ray images of the blood vessel with injection of contrast medium were acquired at 25 fps using a PC frame capture card. Imaging of a Perspex calibration cube allowed 3D reconstruction of the vessel and determination of true dimensions. Blood flow waveforms were calculated off-line on a Sun workstation using the new algorithm. The correlation coefficient between instantaneous blood flow values obtained from the EMF and the x-ray method was r equals 0.871, n equals 1184, p less than 0.0001. The correlation coefficient for average blood flow was r equals 0.898, n equals 16, p less than 0.001. We have successfully demonstrated that our new algorithm can measure pulsatile blood flow in a vessel phantom. We aim to use this algorithm to measure blood flow clinically in patients undergoing vascular interventional procedures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.