Abstract
The tortuous geometry of the human vasculature system, and pulsatile nature of blood flow, combine to generate a complex 3D blood flow pattern. Conventional Doppler ultrasound, however, provides velocity data only in one dimension - in the direction of the ultrasound beam. 3D velocity results are extrapolated form this one dimensional data by assuming the blood is parallel to the vessel axis. A two-dimensional vector Doppler system has been developed to improve hemodynamic visualization, and investigate the possible errors introduced by conventional Doppler methods. The results from a study of a normal human femoral bifurcation demonstrate that flow is not always para-axial. When flow is para-axial, conventional Doppler methods work as expected: the calculated velocity is independent of interrogation angle. When flow is non-axial, however, the velocity waveforms and measurements vary greatly with interrogation angle.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.