Abstract

We have designed and evaluated (using computer simulations) several different 2D phased arrays with square elements for deep localized hyperthermia. These array include a 20x20 planar array, a 80x16 cylindrical-section array, and a 16x16 spherical-section array. Also, we have designed a new phased array with circular elements. We used single and multiple focus scanning methods with intensity gain maximization. We found that the array with circular elements is more effective in reducing grating lobes compared to the same array with square elements. The grating lobes were at least 30 dB smaller than the phased array with square elements and intensity gain was at least 1 dB greater. In addition, with equal intensity distribution patterns for rectangular and circular phased arrays, the number of elements in the circular phased array was smaller and its intensity gain was greater than the other arrays.© (2001) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.