Abstract

The inhomogeneity of tissue structure greatly affects the sensitivity of tissue oxygenation measurement by reflectance NIRS. We have proposed a method for correcting the influence of a subcutaneous fat layer on muscle oxygenation measurement. In this study, this method was validated by measuring the peak-to-peak variation of muscle oxygenation in periodic exercise tests on the vastus lateralis and the falling rate of oxygenation in ischemia tests on the forearm. A newly developed multisensor probe with source- detector distances of 7-40 mm was used. THe probe, consisting of a two-wavelength LED and four photodiodes, was connected to a 4-channel tissue oxygen monitor. The fat layer thickness was also measured by ultrasonography. Results of the tests clearly showed that the presence of a fat layer greatly decreases the sensitivity of measurement and increases the light intensity at a detector. The correction factors of sensitivity were determined from this relationship and Monte Carlo simulation. The corrected oxygenation levels were quantitatively compared among subjects in spite of different fat layer thicknesses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.