Abstract
This paper describes results using a nonabsorbing wavelength as an internal reference to correct for nonanalyte-related signal changes and long-term drift in an absorbance-based fiber optic sensor. A renewable-reagent sensor developed for seawater pCO<SUB>2</SUB> (CO<SUB>2</SUB> partial pressure) was repetitively calibrated over a 12-day period and the drift in the calibration was studied. The results indicate that although the internal reference significantly improves the long-term performance, it does not fully correct for all sources of drift over the 12-day period.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.