Abstract

We conducted visible/near infrared optical measurements on the forearm of human subjects using a commercial diffuse reflectance spectrophotometer, and a breadboard temperature- controlled localized reflectance tissue photometer. Calibration relationships were established between skin reflectance signal and reference blood hemoglobin (Hb) concentration, or hematocrit values (Hct). These were then used to predict Hb and Hct values from optical measurement in a cross validation analysis. Different linear least- squares models for the prediction of Hb and Hct are presented and shows the ability to predict both. It was possible to screen prospective blood donors with low Hb concentration. It was possible to predict anemic subjects in the limited prospective blood donor population.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.