Abstract

A quantitative comparison of the capabilities to produce computerized tomograms was made between the EMI head scanner and reconstructions from images provided by a large screen low light level -TV camera fluoroscopic system. A phantom made from lucite containing rods of various materials and sizes was used. The computer printout of each was analyzed and a high degree of correlation (r = 0.98) was noted between the results of both systems. The differential attenuation detectability of the fluoroscopic system was found to be comparable to or better than the EMI unit. As expected from a consideration of the quantum statistics for each system, the noise in the obtained reconstructions was also comparable. It is concluded that such a fluoroscopic system performs favorably when compared to the presently available commercial systems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.