Abstract

Although the possible interference of digoxin-like immunoreactive substances (DLIS) on the Architect iDigoxin chemiluminiscent microparticle immunoassay (CMIA) has been emphasized by the manufacturer, a specific study about this subject has still not been carried out. Apparent serum digoxin concentrations were determined using the Architect iDigoxin CMIA from Abbott Laboratories in digoxin-free pregnant women (n = 50), and patients with liver disease (n = 50), renal insufficiency (n = 50), kidney (n = 25) or liver (n = 25) transplant, and critical illness (n = 50). In all of the patients included in this study, apparent serum digoxin concentrations were lower than the correspondent quantification limit (< 0.30 microg/L). The Architect iDigoxin CMIA assay would be relatively free from endogenous DLIS positive interferences.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.