Abstract
We evaluated the performance of low-field MRI in breast disorders by comparing it with high-field MRI and biopsy results. Twenty-eight consecutive patients who were able to undergo two magnetic resonance examinations on the following days were examined by high-field and low-field MRI. After T1-weighted sagittal images had been obtained, a dynamic 3-D axial study was performed, followed by the acquisition of contrast-enhanced T1-weighted sagittal images. The images were analyzed separately by two radiologists paying attention to lesion morphology and enhancement kinetics. Six patients had problems in both breasts (34 breasts studied). The results were compared with biopsy results of 27 breasts. There were 16 malignant lesions, 2 fibroadenomas, and 9 other benign lesions. The inter-magnetic-resonance-scanner κ value was 0.77 (substantial agreement), whereas the interobserver κ value was 0.86 and 0.81 at low and high field, respectively (excellent agreement). The sensitivity was 100 and 100%, the specificity was 82 and 73%, and the accuracy was 93 and 89% at low and high field. The mean lesion size was 2 cm and the smallest malignant lesion was 8 mm in diameter. Low-field MRI is a promising tool for breast imaging. Larger materials and smaller lesions are needed to evaluate its true sensitivity and specificity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.