Abstract
A windowless Si(Li) detector was calibrated in the energy range 0.53-14 keV. We used the continuous bremsstrahlung spectra produced by an electron beam on a thick carbon target as a reference. The theoretical shape of the spectra was calculated with the Wentzel formula corrected for attenuation of the soft radiation in a thick target. The continuous bremsstrahlung allows a very precise measurement of the discontinuities in the efficiency curve, which occur due to the x-ray absorption edges of the detector material. The accuracy of this calibration method is limited by the detector's resolution and by layers of ice on the crystal's surface.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.