Abstract

This study develops an improved Feldkamp–Davis–Kress (FDK) reconstruction algorithm using non-local total variation (NLTV) denoising and a cubic B-spline interpolation-based backprojector to enhance the image quality of low-dose cone-beam computed tomography (CBCT). The NLTV objective function is minimized on all log-transformed projections using steepest gradient descent optimization with an adaptive control of the step size to augment the difference between a real structure and noise. The proposed algorithm was evaluated using a phantom data set acquired from a low-dose protocol with lower milliampere-seconds (mAs).The combination of NLTV minimization and cubic B-spline interpolation rendered the enhanced reconstruction images with significantly reduced noise compared to conventional FDK and local total variation with anisotropic penalty. The artifacts were remarkably suppressed in the reconstructed images. Quantitative analysis of reconstruction images using low-dose projections acquired from low mAs showed a contrast-to-noise ratio with spatial resolution comparable to images reconstructed using projections acquired from high mAs. The proposed approach produced the lowest RMSE and the highest correlation. These results indicate that the proposed algorithm enables application of the conventional FDK algorithm for low mAs image reconstruction in low-dose CBCT imaging, thereby eliminating the need for more computationally demanding algorithms. The substantial reductions in radiation exposure associated with the low mAs projection acquisition may facilitate wider practical applications of daily online CBCT imaging.

Highlights

  • This study develops an improved Feldkamp–Davis–Kress (FDK) reconstruction algorithm using non-local total variation (NLTV) denoising and a cubic B-spline interpolation-based backprojector to enhance the image quality of low-dose cone-beam computed tomography (CBCT)

  • The CBCT images resampled with the cubic B-spline interpolation using 16 pixels (4 × 4) were smoother and had fewer interpolation artifacts, resulting in slightly improved reconstruction accuracy for SL, anisotropic total variation (ATV), and NLTV

  • This decreases the number of photons reaching the detector, reducing the signal-to-noise ratio (SNR) of the projection data

Read more

Summary

Introduction

This study develops an improved Feldkamp–Davis–Kress (FDK) reconstruction algorithm using non-local total variation (NLTV) denoising and a cubic B-spline interpolation-based backprojector to enhance the image quality of low-dose cone-beam computed tomography (CBCT). The proposed algorithm was evaluated using a phantom data set acquired from a low-dose protocol with lower milliampere-seconds (mAs).The combination of NLTV minimization and cubic B-spline interpolation rendered the enhanced reconstruction images with significantly reduced noise compared to conventional FDK and local total variation with anisotropic penalty. The most straightforward way to reduce dose of a CBCT system is to use either a lower mAs of the projection data or a shorter exposure time per p­ rojection[9] Utilization of these approaches leads to excessive noise in the reconstructed images. An improved FDK-based CBCT reconstruction algorithm is proposed based on non-local total variation (NLTV) filtration and cubic B-spline interpolation for low mAs of projection data. The effectiveness of the proposed method is demonstrated using phantom studies

Methods
Results
Conclusion