Abstract

In this paper, a rapid and low-cost fabrication technology for high-density microvalve arrays is presented. For proof of the concept, a pneumatically-actuated three-way microvalve system composed of three independent one-way valve units was fabricated and tested. Each valve unit has a membrane, which is actuated by external negative air pressure. Intervals between the valve units are smaller than 780 /spl mu/m. These small intervals have been realized by providing the system with a layer of microchannels to conduct the air pressure to the valve units. All the parts were made of inexpensive silicone elastomer - polydimethylsiloxane (PDMS) - and rapidly fabricated with molding and spin-coating. A newly developed technique for wafer level transfer of a PDMS membrane has been proven to be effective. Flow characteristics of the microvalve system for water are presented. The microvalve works in an on-off manner with hysteresis. No leakage has been observed in the closed state. In the open state, measured flow resistances (pressure drops) are within the range of 1.65-2.29 kPa/(/spl mu/L/min).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.