Abstract

We describe two setups that were built for low-background gamma-ray spectrometry in the Garching Underground Laboratory (∼ 10 m.w.e.). Both setups are based on HPGe detectors surrounded by several layers of passive shielding as well as an active muon veto. The first setup (GEM) comprises a single HPGe detector surrounded by a NaI(Tl) scintillation detector that serves as anti-Compton veto. The second setup (LoAx) consists of two smaller HPGe detectors which are arranged face-to-face to cover a large solid angle around the sample. The detection efficiency of both systems is determined using a calibrated Monte-Carlo simulation. The count rate finally achieved in the energy range 40-2700keV is 10250±26cts/day for the GEM setup, and 5258±27cts/day and 6876±31cts/day between 20-1500keV for the two detectors of the LoAx setup. This leads to detection sensitivities of a few mBq/kg for U and Th at both screening stations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.