Abstract
Purpose: Computed tomography angiography (CTA) is a non-invasive procedure to evaluate vascular anomalies. However, the high radiation dose related to the examination is a concern due to its associated cancer risk. Therefore, this study aims to investigate the effect of low tube voltage CT lower limb angiography with aortogram (CTLLA) on radiation dose and image quality based on body mass index. Methods: Eighty patients were included in this study, where an initial 20 patients were scanned with standard dose protocol, irrespective of BMI. The remaining 60 patients were equally allocated to BMI-based protocol, where high and obese BMI was allocated to 120 kVp protocol, normal BMI to 100 kVp protocol, and low BMI to 80 kVp protocol. Dose data and image quality data were collected after the scan. Results: The result showed a 75% reduction in radiation dose, a 40% increase in vessel attenuation for the 80 kVp protocol, and an improved qualitative image score. Conclusion: In conclusion, tailoring the tube voltage with BMI can be used for CTLLA while achieving significant dose reduction without compromising image quality.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.