Abstract

A method based on the low-temperature electrothermal vaporization of an 8-quinolinolato complex from a tungsten boat cuvette is proposed for introducing a trace amount of aluminum present in a digested steel-making alloy or opal glass sample into an inductively coupled plasma for atomic emission spectrometric detection. The operating conditions were optimized and the effect of foreign ions was investigated. Zirconium(IV) was found to reduce the interferences from other foreign ions. The detection limit was found to be 0.12ng of aluminum and the precision in relative standard deviation for 2.0 ng of aluminum was determined to be 6.1% (n=6). The aluminum contents in some standard samples determined by the proposed method were in good agreement with the certified values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.