Abstract

An exposure facility for wide application to cell exposure to an ELF (extremely low frequency) magnetic field was developed. It is suitable for conducting experiments under a high-intensity, variable-frequency magnetic field, on the biological effects of the ELF magnetic field in an in vitro study. The exposure system consists of Merritt's 4-square coil as a basic component to generate the required magnetic field intensity of 10 mT at 50 Hz with spatial field uniformity less than +/-3% in a 400 mm cube. Concentric compensation coils are adopted to eliminate the effects of stray fields on sham (control) samples in the vicinity of the exposure system. The uniformity of the magnetic field in the exposure coil, the increase in the power supply capacity due to the existence of compensation coils, and the stray field estimation were investigated carefully. After fabricating the system, performance tests were carried out and all the characteristics were found to be satisfactory. In addition, the ideal configuration for a concentric coil system was proposed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.