Abstract

A low-pressure helium microwave-induced plasma was interfaced to a mass spectrometer and the system was utilised as a detector for gas chromatography. The over-all system was optimised with respect to microwave power and the first-stage pressure of the mass spectrometer. The linear dynamic range for iodobenzene extended over three orders of magnitude starting at 1 pg. The absolute 3σ detection limit for iodine was 0.1 pg. The evaluation of the interface for the detection of bromine indicated an interference at mass 79 due to the formation of a polyatomic nickel species. A preliminary detection limit of 3.5 pg was calculated for bromine. A second interface design was used in a preliminary investigation of the backgrounds at masses 35 and 37. The background 35/37 ratio was correct, indicating a source of chlorine within the system. The detection limit for chlorine was 24 pg.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.