Abstract
Several reports have described decreased oxygen extraction fractions in tumours relative to those in normal tissues as measured with the oxygen-15 steady state technique and positron emission tomography. The present simulation study was carried out to assess the influence of tissue heterogeneity on these measured values. It was found that, within the range analyzed, tissue heterogeneity always resulted in underestimations of mean values of oxygen extraction fraction. It must, therefore, be concluded that the oxygen-15 steady state technique is not an accurate method for the assessment of the oxygen status of tumours. This finding should also apply to other pathological conditions, where a significant degree of tissue heterogeneity can not be excluded. More generally, this study demonstrates the need for detailed analyses of sensitivities of tracer kinetic procedures to tissue heterogeneity.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.