Abstract

A liquid cooled, pulsed electromagnet of solenoid configuration suitable for duty in an ultra-low field nuclear magnetic resonance system has been designed, fabricated and successfully operated. The magnet design minimizes Johnson noise, minimizes the hydrogen signal and incorporates minimal metal and no ferromagnetic materials. In addition, an acoustically quiet cooling system permitting 50% duty cycle operation was achieved by designing for single-phase, laminar flow, forced convection cooling. Winding, conductor splicing and epoxy impregnation techniques were successfully developed to produce a coil winding body with integral cooling passageways and adequate structural integrity. Issues of material compatibility, housing, coolant flow system and heat rejection system design will be discussed. Additionally, this pulsed electromagnet design has been extended to produce a boiling liquid cooled version in a paired solenoid configuration suitable for duty in an ultra-low field nuclear magnetic resonance system. This pair of liquid nitrogen cooled coils is currently being tested and commissioned. Issues of material compatibility, thermal insulation, thermal contraction, housing and coolant flow design are discussed.